Home > References > Particle Traps      
Particle Traps

For a customer in the analytical chemistry, Micronit Microfluidics developed wet-etching techniques to create structures for trapping particles and/or cells (see image - figures are in µm). For these structures no additional processes, nor other etching techniques are necessary. The structures can simply be incorporated in the custom made glass chips.


Particle TrapsSpecifications

Chip size at least 5 x 5 mm
Material glass
Technique HF etching
Channel width 10 µm - 600 µm
Channel depth 10 µm and 20 µm
Particles held by trap depending on specs, but > 1.0 µm


Special features:

  • particle trap can be incorporated in channels or in crossing of channels;
  • number of traps can be changed, increasing or decreasing size of trapped particles.


Application
Thanks to the particle traps, for instance, clotting in a mixed channel can be held. The residue can be used for new chemical reactions in the chip. Dependable on the size of the particles the muzzle in the above figure of the particle trap can be made wider, to prevent clogging. The size of the trapping depends on the depth of the channels.


Capabilities

Micronit is capable of making complex particle traps in microfluidic devices of 10 x 10 mm or even smaller. Eventually we can integrate electrodes to measure the particles.